Re: [AMMRL] Logfiles from ICON NMR

From: Fulton, Luke via groups.io <fko24003=uconn.edu_at_groups.io>
Date: Wed, 26 Nov 2025 18:09:51 +0000

Hi Christian,

The Inmracct.brief or Inmracct.full files should have the raw data you're
looking for. I presume the pathway is similar between windows and linux machines,
and adjust for your specific version of TS. C:\Bruker\TopSpin4.4.0\conf\instr\remote_spect\inmrusers
The files list experiments sequentially as they execute. Compiling it all into
more condensed formats requires an external tool. If anyone's curious then
you're welcome to send me a private message and I'll share a generic version
of my excel template.

Kind regards,
Luke


Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
Unit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279

email: fko24003_at_uconn.edu
(alias): luke.fulton_at_uconn.edu
mobile: (603) 953-5275
Office: (860) 486-4069
________________________________
> From: main_at_ammrl.groups.io on behalf of Christian Adam via groups.io
> Sent: Wednesday, November 26, 2025 11:40 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Logfiles from ICON NMR


Dear Spinlanders,

in a current attempt to get hold of the overall instrument usage time, I have
been trying to find the ICON NMR log files. I know that some years ago I
Accessed them for some reason, and they provided - in a very organized manner -
an overview about the samples, the used experiments, and start and stop times
of the experiments, among a whealth ofother information. However, I can't seem
to find these files anymore, and I remember that their location was not really
straightsforward. I know that some of you use these files also for billing
purposes, nmaybe someone can give me a hint where to look in the topspin directory?

Thanks a lot in advande!

Christian


Dr. Christian Adam
Analytical and Material Science; Polymer, heteronuclear and quantitative NMR

Mobile: +49 172 6352576, Fax: +49 621 66-58670, Email: christian.k.adam_at_basf.com
Postal Address: BASF SE, RGS/AS - B009, Carl-Bosch-Strasse 38,
67056 Ludwigshafen am Rhein, Deutschland



BASF SE, Registered Office: 67056 Ludwigshafen, Germany
Registration Court: Amtsgericht Ludwigshafen, Registration No.: HRB 6000
Chairman of the Supervisory Board: Kurt Bock

Board of Executive Directors:

Markus Kamieth, Chairman;
Dirk Elvermann, Michael Heinz, Anup Kothari, Stephan Kothrade, Katja Scharpwinkel


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3010): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3010__;!!PvDODwlR4mBZyAb0!WY4Vd6qZWwZBAXLde3tux8fwBLYMKX=
A7VdXRyWRU5dMeNzfla5kyiBCJEiwdphVO16FaCdjwzd-2OUmo6uDCY5VyA3hL$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116488623=
/7559972__;!!PvDODwlR4mBZyAb0!WY4Vd6qZWwZBAXLde3tux8fwBLYMKXA7VdXRyWRU5dMeN=
zfla5kyiBCJEiwdphVO16FaCdjwzd-2OUmo6uDCY5Yv8hc6$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Nov 26 2025 - 10:39:31 MST

This archive was generated by hypermail 2.4.0 : Mon Dec 01 2025 - 17:39:38 MST