Hello - In case it helps someone else, I was able to get atma working on
this iProbe (at least for now). The "fix" involved removing the probe
from the magnet and issuing an "atma reset" command with the probe on the
floor without a sample.
Many thanks to Thomas Smith for this tip!
Cheers,
Mike
-=-=-=-=-=-=-=-=-=-=-=-