Hi Ken,

Start iconc and go to Lock/Shim options.
Turn off Lock only after a solvent change. That might help.

Best,
Jeff
________________________________
> From: main_at_ammrl.groups.io on behalf of Kenneth Sharp-Knott via groups.io
> Sent: Wednesday, November 5, 2025 3:12 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Incorrectly refereced C6D6

Forgive me, I feel like this is something I should know....

I've been working with a C6D6 sample today and about ½ the time it's
completely incorrectly referenced to the point that ½ of it is off the
scale. The samples are run on Bruker in ICON, yes, I'm selecting the correct
solvent... 🙂 Unfortunately, sometimes it happens during an HSQC
and it gets clipped due to the au program setting the sweep width.


  1.
Why does this happen? I assume the referencing is set somehow during the
lock, so I guess it's locking on the wrong resonance?
  2.
How can I prevent this?

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2976): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2976__;!!PvDODwlR4mBZyAb0!TA3pyDOy8da5XjqT4nGRsWSHpcpL2y=
w58EOiv1p2ek01yCZ-Zky9YFoB_AY_5RmYCqC3nBNoDVQZ7gF0IGZd-TIZLnTj$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116141363=
/7559972__;!!PvDODwlR4mBZyAb0!TA3pyDOy8da5XjqT4nGRsWSHpcpL2yw58EOiv1p2ek01y=
CZ-Zky9YFoB_AY_5RmYCqC3nBNoDVQZ7gF0IGZd-UUyu7rc$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Thu Nov 06 2025 - 07:51:57 MST

This archive was generated by hypermail 2.4.0 : Mon Dec 01 2025 - 17:39:38 MST