Re: [AMMRL] Bruker Topspin 4.2 vs 4.3

From: Scott Burt <scott.burt_at_byu.edu>
Date: Thu, 9 Nov 2023 16:00:48 +0000

  Sorry if this reply is a bit late. Life has been hectic the past few weeks.

When it was first released, I tried 4.3 on our spectrometer (Avance Neo,
RHEL 7) and as a data station (both on RHEL 8 and on Mac OS 12.7). I was
able to operate the spectrometer just fine (granted, the testing was quite
limited). The primary problem that caused me to roll back was that it doesn't
seem to be able to handle symlinks at all (!!).

All of our user data is archived to a network drive and ~/networkdata is
symlinked to the obnoxiously long location on the mounted network drive.
TopSpin would simply not show anything existing within the folder if you
descended into a symlink. On the Mac version, the data server would never
get past the initial synchronize step on any folder that contained a symlink.
Avoiding symlinks is simply not an option with how I have things setup (both
on the user computers and my personal computer).

I've had filing a bug report on my to do list for a while now, but I simply
haven't had the time with my teaching schedule and all the helium recovery
stuff I'm dealing with on top of the NMR lab...

-Scott

Dr. Scott Burt
Professor
NMR Facility Manager
Department of Chemistry and Biochemistry
Brigham Young University
C414 BNSN / C008A BNSN
Provo, Utah 84602-5700
Phone: (801) 422-2404
website: https://chembio.byu.edu/nmr-facility
email: scott.burt_at_byu.edu
________________________________
> From: main_at_ammrl.groups.io on behalf of Kenneth Sharp-Knott via groups.io
> Sent: Monday, October 30, 2023 10:04 AM
> To: ammrl_at_groups.io <ammrl_at_groups.io>
> Subject: [AMMRL] Bruker Topspin 4.2 vs 4.3

I have seen a lot of comments from users who have said they 'rolled back' to
4.2 after numerous issues with 4.3.

I would appreciate it if you could elaborate on what those issues are specifically.
Bruker would like the feedback as they are releasing a new version soon and want
to make sure those issues are taken care of.

One individual stated that they believed there to be more issues with 4.2 than
4.3 so they are surprised by this feedback.

Best

--
Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#773): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/773__;!!PvDODwlR4mBZyAb0!U_QqBGrAmk48xT_hrRcHPmGivwyNjuQq=
kJDozgsIaSOjwO_OarmgZ3IihTmlALTpHP7t7Dsp-Cl1wzL0MJEL$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102278454=
/7559972__;!!PvDODwlR4mBZyAb0!U_QqBGrAmk48xT_hrRcHPmGivwyNjuQqkJDozgsIaSOjw=
O_OarmgZ3IihTmlALTpHP7t7Dsp-Cl1w6amkIWn$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Nov 09 2023 - 08:00:54 MST

This archive was generated by hypermail 2.4.0 : Fri Dec 01 2023 - 12:46:40 MST