Re: AMMRL: Varian 300Mercury SMS Autosampler issues

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 26 Nov 2014 12:38:57 +0100

Dear Lingchao,

> We have Varian 300Mercury SMS Autosampler for our automation runs. I have two issues about this robot now:
>
> 1. When the robot became shaky, the potentiometer inside needs to be replaced. The Zymark potentiometer is now from Perkin Elmer Life (part#: 36816&36817) and it is back ordered until March 2015. Are there any other alternative brands of potentiometer for this SMS autosampler I can use?

I don't have an answer for this (yet) - would be interested in spare parts as well...

> 2. Calibration of SMS Autosampler needs MS-DOS Laptop with 9-pin PCMCIA serial card. It also needs floppy driver in the laptop. (Or Toshiba 1800/1850). Our old laptop seems to be crashing very often so I am trying to buy another laptop for replacement. Does anyone know anywhere still sells those old laptops?

You don't strictly need a laptop with PCMCIA serial card - all you need is any computer that can run plain MS-DOS, has a floppy drive and a DB9-pin serial port. Ebay might have something for you. I guess any 386 (possibly 486, not sure) should be OK for this.

Best regards,
Bert.


Magnetic resonance consulting, service, training & more
Selling & buying used Varian/Agilent NMR equipment
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
 
Spin-Doc
Kiefernweg 13
58239 Schwerte
Germany
Tel. & Fax: +33 367 113463 (French office)
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (www.stelar.it)
MestreLab NMR Software (www.mestrelab.com)
Armar NMR solvents (www.armar.ch)
Received on Wed Nov 26 2014 - 01:39:04 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 18 2023 - 18:20:25 MST