Dear AMMRL,
last summer I posted a query for help on an issue with a Windows XP SP2
host running Xwin 3.5 losing communication to a Avance 500 console.
Initially we focused our attention on hardware fixes (new CCU and etc) then
on networking issues. It eventually became clear that the problem was due
to the Windows Firewall blocking the NFS server connection that Xwin needs
for PC-CCU communication.
Bruker has a script on their website to fix this:
http://www.bruker-biospin.de/NMR/nmrsoftw/passwd/updates/pc/xwinnmr/xwnmr3_5/index.html
or you can set an exception in the Windows Firewall settings for the program:
HCWINSVR.EXE
which is what we did.
In retrospect it was a simple fix but it took us awhile to get there, so
maybe this post will save other people some time.
Jeff
Jeffrey S. de Ropp, Ph.D.
NMR Facility
UC Davis
One Shields Avenue
Davis CA 95616
jsderopp_at_ucdavis.edu
530-752-7677 office
530-752-6480 MS1-D Lab
530-752-5682 600 MHz Lab
530-754-8238 Chem NMR Lab
530-752-3516 Fax
NMR Facility Web site URL is:
http://www.nmr.ucdavis.edu/