Re: [AMMRL] TopSpin installation & CodeMeter issues

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Fri, 29 May 2026 17:25:38 +0000 (UTC)

There are two factors here.  Make sure to use firefox--it's the only
browser I've gotten to work successfully.  And for some reason the license
central webpage has been coming up in offline activation.  If you've got
a link in the lower right that says direct license transfer, click that
and it should do the automatic activation.  

Cheers,Mike

On Friday, May 29, 2026 at 11:20:24 AM MDT, Kenneth Sharp-Knott via groups.io wrote:

I had the same issue.  Bruker told me to use a different browser (I was
using Chrome), but my computer died before I could give it a shot.
They said they were having issues with Chrome.

A student said they had success doing the offline activation.

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry 
Virginia Tech
(540)267-6502 (Cell)(540)231-0885 (Office)

> From: main_at_ammrl.groups.io on behalf of Sameer Al-Abdul-Wahid via groups.io
> Sent: Friday, May 29, 2026 12:32 PM
> To: Main <main_at_ammrl.groups.io>
> Subject: [AMMRL] TopSpin installation & CodeMeter issues

Good afternoon all,

We encounter an occasional issue installing TopSpin on Windows 11 computers.
We can input a license key, but the webpage is not automatically selecting
a CmContainer and there is “No CmContainer found”. I’ve tried
reinstalling CodeMeter from the TopSpin installer and also directly from Wibu to no avail.
There is a Bruker_BBIO_WMLV container at http://localhost:22352/container.html
but it has no licenses. 

This one has me stumped - I’d be grateful for any suggestions!

Best,Sameer 

--
Sameer Al-Abdul-Wahid, Ph.D.
Manager, Nuclear Magnetic Resonance Centre
University of Guelph

519-824-4120 x58914 s.wahid_at_uoguelph.ca



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3495): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3495__;!!PvDODwlR4mBZyAb0!XSj0d2YANN2a_3qRXYFzUwmzf_z8FM=
_upZujauH-xXOk-tb9Qo_bI-9uqo1L6Z1Jb8vRsJ-nQ97WzzJgyV-ywtTEnuJ9$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/119550449=
/7559972__;!!PvDODwlR4mBZyAb0!XSj0d2YANN2a_3qRXYFzUwmzf_z8FM_upZujauH-xXOk-=
tb9Qo_bI-9uqo1L6Z1Jb8vRsJ-nQ97WzzJgyV-ywiXWxABn$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri May 29 2026 - 11:17:51 MST

This archive was generated by hypermail 2.4.0 : Sun May 31 2026 - 13:58:17 MST