Hi Ken
the stator might no have properly flipped. While the probe is outside the
magnet connect the BSB cable and use the "re-flip" option in MAS display ->
configuration. The option is towards the bottom of that window. You should
be able to see if the stator goes properly into the vertical position.
Best regards
Clemens
On 5/26/2026 11:58 AM, Kenneth Sharp-Knott via groups.io wrote:
I have a 4mm SSNMR iProbe with shuttles for sample loading. Recently, a
sample has failed to eject and needs to be manually removed. I tried
inverting the probe, but the sample didn't fall out.
Any tips?
Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech