Re: [AMMRL] problem about shim coil temperature control in two shim system

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Sun, 24 May 2026 06:54:32 +0800
Hi,

A photo of the flush gas adapter could be nice (maybe on a Google drive of
Dropbox?) - but that flush gas adapter should normally be a ring sitting right
on top of the magnet bore and is of utmost importance to protect your magnet from
quenching during (very) low-temperature experiments where the exhaust gas may cool
down the (usually black ring) magnet vacuum flanges and their embedded rubber O-rings
from shrinking and subsequent vacuum loss… so that low shim temp warning is very
serious and usually means your shim flush gas flow is insufficient!

Best regards,
Bert

->Sent via mobile <-

Spin-Doc 
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab software, Stelar relaxometers & Barthel HF amplifiers



Am 23.05.2026 um 23:19 schrieb Sham, Shing via groups.io <ssham=luc.edu_at_groups.io>:

Hi Junfeng,
May not be too helpful here. 
Sorry, I am little confused about your dual system setup.
First, my understanding is that two different shim files are used for the wide-bore shim and standard-bore shim stacks.
Secondly, the wide-bore shim stack is temperature controlled by running water through a separate chiller. There is no flush gas involved in the wide-bore micro-imaging shim stack. The flush gas is mainly used for the probes, particularly for low and high VT in NMR, as far as I know.

Just my 2 cents.

Good luck.

Simon
 



_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Onl= ine (#3485) | Reply to Group | Reply to Sender | Mute This Topic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
Received on Sat May 23 2026 - 16:34:41 MST

This archive was generated by hypermail 2.4.0 : Sun May 31 2026 - 13:58:17 MST