Hello Weixing,
We routinely use Bruker Z172600 NMR tubes
https://store.bruker.com/en-CA/products/unracked-7-nmr-tubes-with-coded-closed-caps
for 600 MHz 1H metabolomics, with excellent results. We regularly achieve a DSS FWHM
between 0.7-0.8 Hz in automation. We have also used a few dozen tubes in VT experiments
at 80 C with no issues.
The Bruker caps do take some getting used to - they are not all that flexible as
compared to the Wilmad caps.
Best regards,
Sameer
--
Sameer Al-Abdul-Wahid, Ph.D.
Manager, Nuclear Magnetic Resonance Centre <
http://nmr.uoguelph.ca/ >
University of Guelph
519-824-4120 x58914 / s.wahid_at_uoguelph.ca<mailto:s.wahid_at_uoguelph.ca>
> From: main_at_ammrl.groups.io on behalf of Zhang, Weixing via groups.io
> Date: Tuesday, May 19, 2026 at 17:41
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] NMR Tube Selection for High-Field Applications
Dear Colleagues,
I would appreciate your input regarding NMR tube selection for high-field
applications.
In the past, I routinely used Wilmad 535-PP-7-5 tubes on a 600 MHz instrument
with good results. However, the current pricing of these tubes has increased
substantially. Both Wilmad and Norell now offer tubes specifically marketed
for 800 MHz applications, such as the Wilmad 541-PP-7-5 and Norell S-5-800-7,
but these products come at a significant premium. In practice, however, I
have not often observed these higher-cost tubes being widely used, even on
800 MHz or 1.1 GHz systems.
From your experience, what are the primary risks or limitations associated
with using lower-specification tubes — for example, tubes nominally rated
for 300 MHz — on a 600 MHz instrument, particularly for quantitative
applications such as determining composition and concentration through signal
integration in different samples?
I would greatly appreciate any guidance regarding acceptable tolerances,
observed deviations, or practical trade-offs.
Best regards,
Weixing Zhang
________________________________
Email Disclaimer: http://www.stjude.org/emaildisclaimer
Consultation Disclaimer: http://www.stjude.org/consultationdisclaimer
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3473): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3473__;!!PvDODwlR4mBZyAb0!X6lT7YW3NqjeLwPl9KDJtdc6COA2YN=
xm3dU4IHb__HWuRtnV89ttoRD5iz_Wj-eqpuDAzRPuKojR9BXa6T55hlCm2MrR$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/119398889=
/7559972__;!!PvDODwlR4mBZyAb0!X6lT7YW3NqjeLwPl9KDJtdc6COA2YNxm3dU4IHb__HWuR=
tnV89ttoRD5iz_Wj-eqpuDAzRPuKojR9BXa6T55hrSYwXm5$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue May 19 2026 - 17:01:21 MST