Re: [AMMRL] 19F peaks issue

From: David Rovnyak via groups.io <drovnyak=bucknell.edu_at_groups.io>
Date: Sat, 9 May 2026 19:07:07 -0400

Signals of interest often lie in this regime so it has to be dealt with
to see those clearly. Unless a special probe is constructed with lower
background, this background can be quite large in typical probes. One
option is that if your native signals are narrow, then a modest amount of
back lp can get rid of most/all of this background without harming your
native lines. -DSR

On Sat, May 9, 2026 at 6:56 PM Eugenio Alvarado via groups.io wrote:

> You can use backward linear prediction to get rid of it. It is likely
> background from the probe.
>
> Eugenio Alvarado
> Director, NMR Lab. Room 3500
> Chemistry Department, University of Michigan
> 930 N. University Ave., Ann Arbor, MI 48109-1055
> Voice: (734)-764-9924
>
> On Sat, May 9, 2026, 5:37 PM Karel Klika via groups.io wrote:
>
>> Hi Puresh,
>>
>> This is just background from the probe, I get the same on my Prodigy (I
>> checked it as it is different to my rt probe). So obviously you have tuned
>> and matched and have a pulse with some power but there is no flourine in
>> your sample.
>>
>> Sincerely,
>> K. Klika
>> ________________________________________
>> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of pothula purushotham reddy via groups.io
>> Sent: 08 May 2026 23:59:10
>> To: main_at_ammrl.groups.io
>> Subject: [AMMRL] 19F peaks issue
>>
>> Hi there,
>>
>> I am measuring ¹⁹F NMR on a Prodigy probe at 400 MHz, but the spectra
>> look strange/weird. Does anyone know what might be causing this or how to
>> fix it? Any suggestions would be greatly appreciated.
>>
>> Thank you!
>>
>>
>> Best
>> Purush
>>

--
---------------------------------------------------------------------------
David Rovnyak, Professor of Chemistry, Bucknell Univ., Lewisburg, PA 17837
BlueSky : NMR_Guy <https://bsky.app/profile/nmrguy.bsky.social >
Research: http://www.facstaff.bucknell.edu/drovnyak/
Bucknell Chemistry: http://www.departments.bucknell.edu/chemistry/
Pascal: "I would like to have written a shorter letter, but I didn't have the time"
--------------------------------------------------------------------------


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3444): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3444__;!!PvDODwlR4mBZyAb0!Uvv2UULH1VyA-ubnXeBTwH7oKvh_vN=
OnuojiuGWvrWntwRaivah4N3AuQ-sDJ8kqjNV8viyTPcEI2eUxUgcyGYs6Ablh$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/119227897=
/7559972__;!!PvDODwlR4mBZyAb0!Uvv2UULH1VyA-ubnXeBTwH7oKvh_vNOnuojiuGWvrWntw=
Raivah4N3AuQ-sDJ8kqjNV8viyTPcEI2eUxUgcyGckPTyWY$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Sun May 10 2026 - 06:18:17 MST

This archive was generated by hypermail 2.4.0 : Sun May 31 2026 - 13:58:17 MST