This is just background from the probe, I get the same on my Prodigy (I checked
it as it is different to my rt probe). So obviously you have tuned and matched
and have a pulse with some power but there is no flourine in your sample.
Sincerely,
K. Klika
________________________________________
> From: main_at_ammrl.groups.io on behalf of pothula purushotham reddy via groups.io
> Sent: 08 May 2026 23:59:10
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] 19F peaks issue
Hi there,
I am measuring ¹⁹F NMR on a Prodigy probe at 400 MHz, but the
spectra look strange/weird. Does anyone know what might be causing this or
how to fix it? Any suggestions would be greatly appreciated.
Thank you!
Best
Purush
-=-=-=-=-=-=-=-=-=-=-=-