I’ve got an 400MHz AVANCE I console that is suddenly giving little / no lock
signal. I had very similar symptoms in April which I tracked down to a loose
2H_TR_J4 cable (jostled during other troubleshooting), but I am seeing identical
issues now which developed suddenly overnight (presumably without students
jostling the console).
Basically, when looking at the lock display, I see very little signal – when
I am unlocked and sweeping with my doped D2O sample, I am see maybe 4x S/N for
the sine waves even using the lock power settings for 90% H2O. Locking is of
course difficult or impossible under these circumstances.
The channel wobbs fine, and a pw90 on 2H yields 400us – slightly higher than
the desired 350, but not so high to explain the problem. Zg2h yields decent
signal.
I went and tightened all the cables on the LCB, R-TX, and L-TX boards. Several
are older and may be a bit frayed – I noticed that this was a problem earlier
this year.
All the lights on the LCB board are green, but the red “error” message
is also on. I can’t find the original AVANCE I LCB manual – I don’t
know what it means specifically.
Is there any chance there’s a repository of technical manuals for the AVANCE
series online somewhere, by the way? I’ve got a scope, but I don’t know
what power levels I should be expecting to detect.
Thanks for any advice,
Beth
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2362): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2362__;!!PvDODwlR4mBZyAb0!Xa63BjWwm_Rbq2my-_MUIsh-KR2SWM=
bz7sbVsfbTjtMD5my5iruddyneBP-IVrgYbF373Jek2ovLnbmwWYg1QGKt8LuZ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113251261=
/7559972__;!!PvDODwlR4mBZyAb0!Xa63BjWwm_Rbq2my-_MUIsh-KR2SWMbz7sbVsfbTjtMD5=
my5iruddyneBP-IVrgYbF373Jek2ovLnbmwWYg1QCUMu5L3$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu May 22 2025 - 10:14:36 MST