Re: [AMMRL] Lost connection to Bruker Cryoplatform #Hardware #Troubleshooting #Bruker #ColdProbes

From: Andrew Fowler via groups.io <afowler13=gmail.com_at_groups.io>
Date: Wed, 14 May 2025 19:04:47 -0400

Since others are asking, here's a copy of what I sent to Greg.

Make sure the crco services are running ok. I'd run the crco_stop then
crco_start scripts. I have had a system lose it's mind and you might have
to do a remove and install.

Under Linux I think the scripts are in /usr/diskless/crco_data (possibly
/prog/bin under diskless, I don't have a TS3.x system handy but it's
definitely under diskless). They should ideally be run as root.

If you're having the same problem under Windows, the scripts are also under
diskless but start with w_ and are in a subfolder labeled something like
win_app, I'm not at my laptop or a console. Make sure to run as a full
administrator.

Cheers,

Andrew

On Wed, May 14, 2025, 9:36 AM Matthew Devany via groups.io wrote:

> Hello Greg,
>
> Could you tell me a little more about how you resolved this? I'm having
> the same issue. I can connect to the crco through UniTool but not
> cryopanel. What are the services and did you roll back your kernel or
> add modifications?
>
> Matthew
>
> On 5/13/25 17:16, gregory.wylie via groups.io wrote:
> >
> >
> > ------------------------------------------------------------------------
> > Hello Everyone
> > Thank you for all your suggestions. We think the recent kernal update
> > took out the crco services. We checked this on Andrew Fowlers
> > suggestion. We have 2 older Bruker systems and we have had several
> > issues after kernal updates with various services being stopped,
> > disabled etc!
> > This did get the cryopanel to open for us but it seems we have a bad
> > turbo as well! Oh the joys!
> > Thanks again everyone!
> > Greg
> > Gregory P. Wylie, Ph.D
> > NMR Facility Manager
> > Texas A&M University
> > Department of Chemistry #3255
> > 580 Ross St.
> > College Station, TX 77843-3255
> > gpwylie_at_tamu.edu
> > 979.458.0705 (voice)
> > 979.845.4719 (fax)
> > 706.206.0007 (cell)
> > http://nmr.tamu.edu
> > <
> http://nmr.tamu.edu/
> >
> >
> --
> *Matthew Devany, Ph.D.*
> Director NMR Laboratory -- Department of Chemistry, Hunter College of
> the City University of New York
> 695 Park Ave., Rm. N1302 -- New York, NY 10065
> Lab: 212-772-5337 -- Fax: 212-396-6327 --
>
> https://hunter.cuny.edu/artsci/chemistry/research/nuclear-magnetic-resonance/
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2347): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2347__;!!PvDODwlR4mBZyAb0!VPvG6HtH2CGPTi6upm4u3QALxh2dB2=
KY60PIaByb7rIQl-O-WjPUPpq4q7AvBnkFbis1DZMyfevuhYmpikZ866iCzUMe$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113090694=
/7559972__;!!PvDODwlR4mBZyAb0!VPvG6HtH2CGPTi6upm4u3QALxh2dB2KY60PIaByb7rIQl=
-O-WjPUPpq4q7AvBnkFbis1DZMyfevuhYmpikZ86-wb24fy$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!VPvG6HtH2CGPTi6upm4u3QALxh2dB2KY60=
PIaByb7rIQl-O-WjPUPpq4q7AvBnkFbis1DZMyfevuhYmpikZ86z-OYamn$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!VPvG6HtH2CGPTi6upm4u=
3QALxh2dB2KY60PIaByb7rIQl-O-WjPUPpq4q7AvBnkFbis1DZMyfevuhYmpikZ86xvpQkVt$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!VPvG6HtH2CGPTi6upm4u3QALxh2dB2KY60PIaB=
yb7rIQl-O-WjPUPpq4q7AvBnkFbis1DZMyfevuhYmpikZ8630A9ROB$
Mute #coldprobes:https://urldefense.com/v3/__https://ammrl.groups.io/g/main=
/mutehashtag/coldprobes__;!!PvDODwlR4mBZyAb0!VPvG6HtH2CGPTi6upm4u3QALxh2dB2=
KY60PIaByb7rIQl-O-WjPUPpq4q7AvBnkFbis1DZMyfevuhYmpikZ86-pnxd8P$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon May 19 2025 - 07:52:26 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 01 2025 - 15:26:23 MST