As Bert says, you may need to change the field
frequency in the cf. What is the field value in the last spectra you recorded
(this value appears in the shim file in the experiment folder)? If the value is
close to -9999 or 9999, you've reached the field limit and should lower or raise
the field frequency by running a new CF.

Best regards.