Re: [AMMRL] Control of nmrsu access in IconNMR

From: T. Andrew Mobley via groups.io <mobleyt=grinnell.edu_at_groups.io>
Date: Fri, 24 May 2024 19:05:04 +0000

Hi all,

Clemens Anklin helped me understand what was going on. To resolve the behavior
it is necessary to set the “Check Passwords” setting in IconNMR
options to “Everywhere” instead of “Sensitive tasks only”.
Now both the nmrsu and nmr passwords are also checked.

Thanks again Clemens,

Andy

T. Andrew Mobley

Personal Pronouns: He/Him/His
For Information on Why I share my pronouns:
http://magazine.grinnell.edu/sites/default/files/documents/Using*20Pronouns_0.pdf

Virtual Office: https://grinnellcollege.webex.com/meet/mobleyt
Making an Appointment: https://calendly.com/mobleyt/30min

Professor of Chemistry
Noyce Science Center 2606
1116 8th Avenue
Grinnell College
Grinnell, IA 50112
641-269-4543


> From: Mobley, Thomas (Andy) <MOBLEYT_at_Grinnell.EDU>
> Date: Friday, May 24, 2024 at 6:07 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Control of nmrsu access in IconNMR

Hi Bob,

I checked, and the nmrsu account on the computer is password protected (I
was 99% sure it was, but I checked anyway). I also checked to make sure that
it was logged out (and not sitting idly in the background) since you can now
do multiple logins on windows (this is a Windows TopSpin installation). The
behavior didn’t change though. I still see the same thing. When I
move back to our standard account (which is an additional user in IconNMR),
it asks me for the password.

So somehow, I think that you are right in thinking that the difference lies
in whether or not it is an account associated with the OS or with the IconNMR
software. I checked the nmr user in IconNMR, and it has exactly the same
behavior.

Andy

T. Andrew Mobley

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1288): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1288__;!!PvDODwlR4mBZyAb0!TZrn6qDS2tuDLJrAUzRSldp9vaIke-=
Cg_kCRt0bh3y7EomGB9lhJGWPQ-OOj3j6R5TBiDMc0XcUoF4eAuwUfkOt1ECRs$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106280679=
/7559972__;!!PvDODwlR4mBZyAb0!TZrn6qDS2tuDLJrAUzRSldp9vaIke-Cg_kCRt0bh3y7Eo=
mGB9lhJGWPQ-OOj3j6R5TBiDMc0XcUoF4eAuwUfkA89d8QL$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri May 24 2024 - 12:05:11 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST