Re: [AMMRL] Assistance Needed: 29Si Experiment Issues #Bruker

From: Craig Grimmer <craig.grimmer_at_gmail.com>
Date: Tue, 14 May 2024 13:52:18 +0200

Good afternoon Muhammad

On issue 2, try increasing the number of scans because it looks like there
is the beginning of the "missing" signals in your spectrum. As a rule of
thumb, 2 x S/N requires 4 x NS.

Kind regards,
Craig.


On Tue, 14 May 2024 at 10:50, Muhammad Adil Raees via groups.io wrote:

> Hello everyone,
>
> We recently conducted a 29Si experiment on our system and encountered two
> issues. We would appreciate your advice on these matters.
>
> Issue 1: I was able to tune and match the probe with ATMM as shown in the
> attached picture. I stored the values by clicking on the close and store
> values box. However, when I tried to perform ATMA, I was directed to
> optimize from ATMM again. A message box popped up on the screen, as
> attached in this email. I am wondering how to store the wobble curve so
> that ATMA can adjust it itself.
>
> Issue 2: The researcher expected a heptate from 29Si at around 188ppm, but
> I only detected a pentate. The signal is clear, but the outermost signals
> didn't appear. I am wondering what parameters I should focus on to improve
> the result. Do you think an inaccurate 90-degree pulse could be the cause?
>
>
> Kind regards,
>
> --
> Muhammad Adil Raees, Ph.D.
> Manager NMR Facility
> The Central Laboratory
> Syed Babar Ali School of Sciences and Engineering
> Lahore University of Management Sciences
> Opposite Sector-U, D.H.A. 54792, Lahore
> +92-330-2866755
> adil.raees_at_lums.edu.pk
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1238): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1238__;!!PvDODwlR4mBZyAb0!RrK-E35MWjohJHBIW83zU25tvzXd_m=
LacrFV7etCINae2Yur7jBjsBioUxihJVA8os94B-5aynnhDgA_zUqkDQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106091125=
/7559972__;!!PvDODwlR4mBZyAb0!RrK-E35MWjohJHBIW83zU25tvzXd_mLacrFV7etCINae2=
Yur7jBjsBioUxihJVA8os94B-5aynnhDgC2bMlFIg$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!RrK-E35MWjohJHBIW83zU25tvzXd_mLacrFV7e=
tCINae2Yur7jBjsBioUxihJVA8os94B-5aynnhDgAhPHvFMA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






29Si.png
(image/png attachment: 29Si.png)

29Si_2.png
(image/png attachment: 29Si_2.png)

Received on Tue May 14 2024 - 05:10:50 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST