Hi All,
This is what worked for me. I’ve only done it once as I am waiting
for additional new workstations to arrive. But I have TS3.7 and Alma working
on an AVIII.
First, if you re-install diskless, the IPSO should boot as that starts the
tftp process. Is that true for you? BUT, on reboot of the workstation, the
process is not started and the IPSO will not boot.
Give command as root or sudo:
systemctl enable tftp
and then reboot the workstation. The IPSO should boot. This is what I have
in my notes and what I am sure worked for me. Now that I look at the system,
tftp is not active so IDK. LMK if this works.
Also for the dhcpd process to start on boot, the ethernet cable must be plugged
into the workstation port and be active. dhcpd is also required for the IPSO to boot.
For the sticky bit issue, I put this in the cron for root:
_at_reboot /usr/bin/chmod -t /tmp
John
> On May 3, 2024, at 8:24 AM, Parnham, Stu <stitch_at_email.unc.edu> wrote:
>
> Hi,
>
> I wish I had information about that. It seems there are a number of us
> that cannot connect to the IPSO when upgrading to Topspin3.7 and Alma9
> Linux. If I hear anything back from Bruker I will post the response.
>
> Stu
>
> -------------------
> Stuart Parnham Ph.D.
> Biomolecular NMR
> Marsico Hall, SB205A
> University of North Carolina
> Phone:(919)-966-6184
> From: main_at_ammrl.groups.io On Behalf Of Craig Grimmer via groups.io
> Sent: Friday, May 3, 2024 8:10 AM
> To: main_at_ammrl.groups.io <mailto:main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Topspin3.7 issues
>
> Hi All
>
> I'd also like that information, please. I have a new workstation (Alma-9.2,
> TopSpin-3.7.0) that fails to "cf" with the error "No AV3 spectrometer detected
> in the spectrometer subnet" even though I can open the web services of the IPSO,
> DRU, AQS, BSMS via a web browser. I've just checked and I see that xinet and
> tftp are not installed.
>
> Thank-you,
> Craig.
>
John Decatur, Ph. D.
Senior Research Scientist, Director of Chemical Instrumentation
and Precision Biomolecular Characterization Facility
Department of Chemistry
3000 Broadway, Mailcode 3179
Columbia University
New York, NY 10027
212-854-2155 (voice)
212-932-1289 (fax)
jdd13_at_columbia.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1219): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1219__;!!PvDODwlR4mBZyAb0!UVFM3reVEZZbWTuhyaI4NXoc-wYDMo=
Dv--wCSI-ndtkXu8MmePSe39BvG0GZjCZjVZdISChdgYGfWp9L$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105862048=
/7559972__;!!PvDODwlR4mBZyAb0!UVFM3reVEZZbWTuhyaI4NXoc-wYDMoDv--wCSI-ndtkXu=
8MmePSe39BvG0GZjCZjVZdISChdgZeZN1CN$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri May 03 2024 - 06:00:49 MST