[AMMRL] AVII on VMbox: Sporadic loss of connection to spectrometer

From: Andre Dallmann via groups.io <andre.dallmann=chemie.hu-berlin.de_at_groups.io>
Date: Fri, 20 Mar 2026 10:24:19 +0100

Dear all,

after a lot of struggle and thanks in part to the brilliant tutorial by
Sameer  (thanks again!!!) for setting up Topspin on VMWare we finally
have our AvanceII console running with Topspin 3.2 on Centos5 running on
Oracle VirtualBox 7.2.6 running on a computer with AlmaLinux9.

However, sporadically, the console completely loses connection to the
computer.

We can either restart both the console and the computer and then
everything works for a while (roughly 8 days). Or we just pull the
ethernet cable going to the console, after that we have the connection
reestablished but not everything in the console can cimmunicate. So we
have to reboot via telnet spect and init 6 and then everything works
again (in our hands only for 24h though).

Does anyone have an idea what the problem might be?

Any suggestion would be greatly appreciated and best regards, Andre

PS: Nobody ever had problems with superhuge icons in TS2.1 running on
VMWare???

--
Dr. André Dallmann
Leiter der NMR-Abteilung

Email: andre.dallmann_at_chemie.hu-berlin.de
Raum: 0'103
Telefon: 82952
Handy: 01797657855
ORCID-ID: https://orcid.org/0000-0002-0141-2635
Webseite: https://www.chemie.hu-berlin.de/en/forschung-en/nmr-spectroscopy/nmr-spectroscopy?set_language=en

Adresse:
Humboldt Universität zu Berlin
Institut für Chemie
AG NMR
Brook-Taylor-Str. 2
12489 Berlin
Deutschland



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3306): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3306__;!!PvDODwlR4mBZyAb0!QfmWiM_DvYh-PFZi8tvUJfU3A7Tsak=
16XMZ8BKQ1nFiGxtmyNUbwnSU9-DI1CDK_XBZGuDn6OCVA_fZ8jQ1bds1Q-gJl$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118415324=
/7559972__;!!PvDODwlR4mBZyAb0!QfmWiM_DvYh-PFZi8tvUJfU3A7Tsak16XMZ8BKQ1nFiGx=
tmyNUbwnSU9-DI1CDK_XBZGuDn6OCVA_fZ8jQ1bdgufIdDh$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Fri Mar 20 2026 - 03:51:39 MST

This archive was generated by hypermail 2.4.0 : Wed Apr 01 2026 - 17:02:32 MST