Re: [AMMRL] : Position of GP1 in HSQCETGPSP

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Wed, 11 Mar 2026 18:35:57 +0000

HI Weixing,

this looks like radiation damping returns water back to +Z towards the end
of the experiment when d0 gets really long. Flipping the sequence to have
the gradient first dephases the water magnetization.

You should also be able to put a gradient between the 90 1H and 90 13C pulses.
At that point magnetization of interest should be IzSz and water is Iy a
gradient at this time will dephase it.

These high-Q probes give you great sensitivity but radiation damping is
something you have to take into account.

Clemens

On 3/11/2026 1:30 PM, Zhang, Weixing via groups.io wrote:

Dear All,

I am running an HSQC experiment of glucose in 90% H₂O / 10% D₂O
using a long d₁ delay.

During acquisition, the FID became dominated by noise toward the end. After
I inverted gp1 and placed it immediately after the first ¹³C 90° pulse,
this issue disappeared and the water suppression improved noticeably.

However, AI-based software tools do not recommend this modification.
I have not observed any differences in the resulting spectra, but I would
appreciate any comments or insights you may have on this adjustment.

Best regards,

Weixing

________________________________

Email Disclaimer: http://www.stjude.org/emaildisclaimer
Consultation Disclaimer: http://www.stjude.org/consultationdisclaimer


--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3290): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3290__;!!PvDODwlR4mBZyAb0!UEyygQ6twCaQTZMfNyjB9JoDOGJMN7=
m5k3lFJtSnaanqkjL_SEVKjjfpce7-NAj8dbHEJtDwDuhcaqdTc57OAy2wg9Op$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118264668=
/7559972__;!!PvDODwlR4mBZyAb0!UEyygQ6twCaQTZMfNyjB9JoDOGJMN7m5k3lFJtSnaanqk=
jL_SEVKjjfpce7-NAj8dbHEJtDwDuhcaqdTc57OAzZrPpmF$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Mar 11 2026 - 11:38:03 MST

This archive was generated by hypermail 2.4.0 : Wed Apr 01 2026 - 17:02:32 MST