Dear Anne

This parity error from the GAB/2 is common on restarting the acquisition
system after a power failure. Don't worry too much about it. If your
GAB/2 gets stuck in an error state on startup, something that happens
occasionally, you can check the status and reset it via the "Gradient..."
link on the BSMS web page.

Kind regards,
Craig.


On Sat, 7 Mar 2026 at 20:59, Anne Rachupka via groups.io wrote:

> Happy Saturday,
>
> My 500 Avance 3 crashed out last night and threw out an error about the *GAB/2
> Partly Error on GCON Interface*. The board had no error lights and
> everything looked fine. I ended up power cycling the console before things
> began responding normally. I’m just curious what that error means– never
> seen it before. Should I be worried about the GAB/2 board? I don’t have
> another one handy to swap in.
>
> Thanks!
>
> _______________________
>
> Anne Rachupka
>
> NMR Director
> University of New Mexico, Albuquerque, NM
> Dept. of Chemistry and Chemical Biology
> 505-277-4031
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3273): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3273__;!!PvDODwlR4mBZyAb0!Vgse3EqVGdISlkoc2haUroiIBh1S69=
OAHgk9g7-37ofONzDM5BjlKcLv71xTS4TtO8C10qnqwZ_DJhOSMtsWXW6Tn6qV$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118193197=
/7559972__;!!PvDODwlR4mBZyAb0!Vgse3EqVGdISlkoc2haUroiIBh1S69OAHgk9g7-37ofON=
zDM5BjlKcLv71xTS4TtO8C10qnqwZ_DJhOSMtsWXTXMn7le$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Sun Mar 08 2026 - 04:49:30 MST

This archive was generated by hypermail 2.4.0 : Wed Apr 01 2026 - 17:02:32 MST