;hoesygpph_2.1
;avance-version (13/08/06)
;2D heteronuclear NOE
;phase sensitive
;
;C. Yu & G. Levy, J. Am. Chem. Soc. 106, 6533 (1984)
;
;$CLASS=HighRes
;$DIM=2D
;$TYPE=
;$SUBTYPE=
;$COMMENT=


#include <Avance.incl>
#include <Delay.incl>
#include <Grad.incl>


"p2=p1*2"
"p4=p3*2"
"d12=20u"


"d0=3u"

"in0=inf1/2"


"DELTA=d0*2+p2"

"TAU=d8-p16-d16-54u"


1 ze
2 d1 do:f2 
3 d12 pl2:f2
  (p3 ph1):f2
  DELTA
  (p4 ph5):f2
  d0
  p2 ph3
  d0
  (p3 ph2):f2
  50u UNBLKGRAD
  TAU pl12:f2
  p16:gp1
  d16
  4u BLKGRAD
  p1 ph4
  go=2 ph31 cpd2:f2
  d1 do:f2 mc #0 to 2 F1PH(ip1 & ip5, id0)
exit


ph1=0
ph2=0 2
ph3=0 0 2 2
ph4=0 0 0 0 1 1 1 1 2 2 2 2 3 3 3 3
ph5=1
ph31=0 2 0 2 1 3 1 3 2 0 2 0 3 1 3 1


;pl1 : f1 channel - power level for pulse (default)
;pl2 : f2 channel - power level for pulse (default)
;pl12: f2 channel - power level for CPD/BB decoupling
;p1 : f1 channel -  90 degree high power pulse
;p2 : f1 channel - 180 degree high power pulse
;p3 : f2 channel -  90 degree high power pulse
;p4 : f2 channel - 180 degree high power pulse
;p16: homospoil/gradient pulse                       [1 msec]
;d0 : incremented delay (2D)                         [3 usec]
;d1 : relaxation delay; 1-5 * T1
;d8 : mixing time
;d12: delay for power switching                      [20 usec]
;d16: delay for homospoil/gradient recovery
;inf1: 1/SW = 2 * DW
;in0: 1/(2 * SW) = DW
;nd0: 2
;ns: 8 * n
;ds: 32
;td1: number of experiments
;FnMODE: States-TPPI, TPPI, States or QSEQ
;cpd2: decoupling according to sequence defined by cpdprg2
;pcpd2: f2 channel - 90 degree pulse for decoupling sequence


;for z-only gradients:
;gpz1: 30%

;use gradient files:
;gpnam1: SMSQ10.100



;$Id: hoesygpph,v 1.1.4.2 2013/08/30 09:44:51 ber Exp $