Re: [AMMRL] can get signal under nmrsu but not under a different user

From: Frank Holger Foersterling via groups.io <holger=uwm.edu_at_groups.io>
Date: Wed, 26 Mar 2025 15:05:09 -0500

Brian,

did you change the base frequency of the spectrometer recently and were
cloning/re-using old data sets? Because that is one case in my
experience where edasp and save without making any change will get the
frequencies in the dataset reset. But that should not effect automation
or reading new parameter files. Those need to be converted with the
expinstall/paracon command executed after the change, but that command
will not change existing datasets.

Holger

On 3/26/25 1:37 PM, Brian Antalek via groups.io wrote:

> I tried  edasp  and found the routing to be as it should.  I then
> resaved it before closing.
> It now works!
> I wonder if it's related to a recent console shutdown and startup?
> thanks for the help!
> Brian
>

--
Frank H. Foersterling
Instrumentation Specialist
Chemistry 362
Department of Chemistry
University of Wisconsin-Milwaukee
414-229-5035
holger_at_uwm.edu



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2173): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2173__;!!PvDODwlR4mBZyAb0!WfapXPaU83gscVByN08W9oZ941wAgV=
OmnENT1MSlVvjlZNfdKCdHWlYew8CKkeMl49wdq1O3awf7YFiCfll7ts0qcw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111903971=
/7559972__;!!PvDODwlR4mBZyAb0!WfapXPaU83gscVByN08W9oZ941wAgVOmnENT1MSlVvjlZ=
NfdKCdHWlYew8CKkeMl49wdq1O3awf7YFiCflk579MH9A$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Wed Mar 26 2025 - 13:15:34 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST