Re: [AMMRL] Recirculating chiller options for use with High Gradient Diffusion NMR probes?

From: Jeff Walton via groups.io <jhwalton=ucdavis.edu_at_groups.io>
Date: Wed, 27 Mar 2024 01:54:43 +0000

Hi Ken,

Replaced ours a long time ago with PolyScience chillers. They have been more
reliable. I haven’t had to replace a pump yet.

Best,

Jeff

On Mar 26, 2024, at 8:16AM, Kenneth Sharp-Knott via groups.io wrote:

Our variable flow refrigerated recirculating chiller (Thermo Neslab Merlin
M33) has died. Long live the chiller. This chiller circulated water through
the gradient coils in our Diffusion probes.

For those of you using a similar high power gradient setup, what are you using
for a chiller? I think the Merlin M33 may have been oversized because it was
necessary to use it at a lower flow rate than it wanted - Low Flow error beeps.
It had a variable flow rate recirculating pump system.

Best

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1086): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1086__;!!PvDODwlR4mBZyAb0!TBNckmnGEFxyx9Yha2_gGM6B6UADGp=
6P313xg-HkPpY0dx-1igoNm7puaHyptVwOI--hr7p9zOBPbpLV5UUfVwu23FtO$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105159955=
/7559972__;!!PvDODwlR4mBZyAb0!TBNckmnGEFxyx9Yha2_gGM6B6UADGp6P313xg-HkPpY0d=
x-1igoNm7puaHyptVwOI--hr7p9zOBPbpLV5UUfV12EaL-Z$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Mar 26 2024 - 18:54:56 MST

This archive was generated by hypermail 2.4.0 : Fri Apr 12 2024 - 18:03:18 MST