Re: [External Email] Re: [AMMRL] Bruker SampleCase: Samples getting stuck as they are dropped down the transport tube

From: Juergen Schulte via groups.io <schulte=binghamton.edu_at_groups.io>
Date: Wed, 06 Mar 2024 10:01:50 -0800

Considering the low pressure, maybe the slider gets stuck because the horizontal
rail needs to be greased again.

That worked for us when we had a similar problem last year (even though we
have 6 bar).

Best Regards,

Jürgen

_______________________________________________

Jürgen Schulte, Ph. D.

NMR Specialist
Binghamton University
SN Building, RM 1024
25 Murray Hill Road
Vestal, NY 13850
 

Ph.: 607-777-4405
Fax: 607-777-4478

schulte_at_binghamton.edu

http://nmr.binghamton.edu/Schulte.html

> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Brown, Silas
> Sent: Wednesday, March 6, 2024 12:42 PM
> To: main_at_ammrl.groups.io
> Subject: [External Email] Re: [AMMRL] Bruker SampleCase: Samples getting stuck as they are dropped down the transport tube

Hello Kenneth,

I do not know if this would be applicable. But when I was at York College of
PA, the compressor we were using would only barely produce the 80psi (5.5 bar)
and I noticed that suddenly one time we kept having an issue with the sample
getting to above the magnet but not releasing. It turned out that the pressure
was enough to get it to the top and start moving but not enough to get it to
the end and let the sample go. It turned out that the issues corresponded with
a large number of people needing the house air at the same time, and that drop
of a few psi was enough to cause the problem.

I don't know if you have a dedicated air supply or a house one but thought
I would add my two cents. (am still pretty new here)


Sincerely,

Silas

--

Silas Brown (He/Him), Instrumentation Specialist
Department of Chemistry
Williams College
150B Morley Science Laboratories
Williamstown, Ma. 01267
E: seb14_at_Williams.edu <mailto:seb14_at_Williams.edu> | P: 413.597.3544





-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1064): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1064__;!!PvDODwlR4mBZyAb0!WouSnjctRtZp--hlyGvT3LODRe8Fyv=
8QsK8ven5_rte9zLPPhnrVNYsDf2OFWnLjznVJBU8WisxyYkDvsO377UqIf_5Q$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104771284=
/7559972__;!!PvDODwlR4mBZyAb0!WouSnjctRtZp--hlyGvT3LODRe8Fyv8QsK8ven5_rte9z=
LPPhnrVNYsDf2OFWnLjznVJBU8WisxyYkDvsO377cELQ-ZT$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Mar 06 2024 - 10:01:53 MST

This archive was generated by hypermail 2.4.0 : Fri Apr 12 2024 - 18:03:18 MST