Re: [AMMRL] Bruker SampleCase: Samples getting stuck as they are dropped down the transport tube

From: ter Horst, Marc A. <terhorst_at_email.unc.edu>
Date: Wed, 06 Mar 2024 07:03:23 -0800

Hi Ken and Hasan

   Yes. On two of our systems with a SampleXpress. Went through the same
adjustments, o-rings, spinners, source compressed gas pressure, sample lift
adjustment. Also tried re-adjusting the bed of the SampleXpress and with
a level. Does this happen with ‘regular’ 5mm tubes or
J-young tubes or both?

  Will try Hasan’s recommendation (Thank you!).

Marc


> From: main_at_ammrl.groups.io On Behalf Of Hasan Celik via groups.io
> Sent: Wednesday, March 6, 2024 10:00 AM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Bruker SampleCase: Samples getting stuck as they are dropped down the transport tube

Hello Ken,

We had this problem before. It turned out that the bolts holding the metal
part that guides the spinner into the transport tube got loose over time
(~ 5 years after the installation). I had to remove the plastic transparent
plastic cover around the unit to access them. Once I lined up the metal part
that latches onto the plastic tube and tightened the bolts, it didn't happen
again.

Best,

Hasan

On Wed, Mar 6, 2024 at 6:54 AM Kenneth Sharp-Knott wrote:

Has anyone had any issues with their spinners getting stuck at the top of
the transport tube as they are dropped? It is a little difficult to see
just why they're getting stuck, but they appear to be kinked at an angle
at the top of the transport tube.

I've changed the rubber band - no effect.
The air pressure may be a bit lower than what Bruker wants 5.5 bar vs 6,
but it's the same exact air pressure I've had since it was installed.

Seeking the usual wise advice and counsel for the AMMRL greats.

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


--
_____________________________
Director, PMRC Core NMR Facility
College of Chemistry
University of California, Berkeley
227A Hildebrand Hall
Office: (510) 643-9829
Cell: (949) 732-8175


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1056): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1056__;!!PvDODwlR4mBZyAb0!UQvCxJDfB4VRMrviR9Uu3L3KvZQ-5G=
Y5Bs2Z80R7QcKUZ8MCO3IeNt61ksjKDGFJ3aJUtUtpXx0DRTI_GxMdEA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104766890=
/7559972__;!!PvDODwlR4mBZyAb0!UQvCxJDfB4VRMrviR9Uu3L3KvZQ-5GY5Bs2Z80R7QcKUZ=
8MCO3IeNt61ksjKDGFJ3aJUtUtpXx0DRTJNiaBnaw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Mar 06 2024 - 07:03:26 MST

This archive was generated by hypermail 2.4.0 : Fri Apr 12 2024 - 18:03:18 MST