Re: AMMRL: cf not working after re installation -SOLVED

From: celine schneider <cmschnei_at_mun.ca>
Date: Thu, 7 Mar 2019 11:37:29 -0330

Thank you all for your help once again! I received many answers, which I’ll summarize below, but what did it for us was to reboot Diskless separately! Go figure…

I should mentioned we had checked many of the other possible causes (network, configuration, permissions, etc…) and all were fine on our end.

-Working solution for us: Turn off AQX, re-install Diskless from the CD (costumer install, diskless only selected). Restart computer, reboot AQX.

Other things to check:

 Unplug the ethernet from the "World", Then try to boot things. This worked for us and may work for you. It could be SELinux, if you are running Linux or your antivirus software

reinstall the Topspin NT toolkit as well
 use the UNITOOL to connect to the console first and just hit enter on all the prompts until you get to the menu. After that close it and try loading topspin again. That seemed to get the startd demon to activate.
Looking at the boot process on tty port will definitively tells you if the CCU booted.
The bootparam and bfs services have to be started and configured well for the CCU to boot correctly.
Hummingbird NFS server that is not configured and running correctly
Is the firewall setting of XP correct to allow the connection to CCU-diskless?

Celine



 
> On Mar 5, 2019, at 5:44 PM, celine schneider <cmschnei_at_mun.ca> wrote:
>
>
> Hi,
>
> Our computer crashed and we had to re-install TS1.3 on a “new” box (running XP because our 500 is AV….). Installation went ok, I had a NMR_backup somewhere (used NMR_SAVE), no problem with Flexlm/license, but when doing cf, I’m getting the following:
>
> "connection to spect (aqport0) failed
> unable to connect to remote host *Connection refused)
> ('startd' demon active on CCU ?)”
>
> I rebooted the AQS, then the entire console, error is still there (we did that at least 10 times).
>
> CCU is flashing nicely.
> I can ping to CCU but I cannot telnet to CCU (it opens and shuts down immediately).
> SPEC is in automatic debit (100MBps). If I switched to 10MBps, ping to CCU times out.
> All services are started as expected.
>
> Troubleshooting CCU mentioned plugging an Hyperterminal in TTY000, but I cannot find it. We have TTY001-TTY020. no 000.
>
> Any suggestion?
>
> Thank you,
>
> Celine
>
> Celine Schneider, PhD
> Adjunct Professor in Chemistry
> Research Laboratory Associate
> NMR Facility / C-CART
> CREAIT Network
> Memorial University of Newfoundland
> St. John's, Newfoundland, A1B 3X7
>
> Tel: (709)-864-3134 <tel:(709)-864-3134>
> email: cmschnei_at_mun.ca <mailto:cmschnei_at_mun.ca>
>
> http://www.mun.ca/creait/CCART/ <http://www.mun.ca/creait/CCART/>
> Sent from my iPhone

Celine Schneider, PhD
Adjunct Professor in Chemistry
Research Laboratory Associate
NMR Facility / C-CART
CREAIT Network
Memorial University of Newfoundland
St. John's, Newfoundland, A1B 3X7

Tel: (709)-864-3134
email: cmschnei_at_mun.ca

http://www.mun.ca/creait/CCART/
Received on Thu Mar 07 2019 - 05:08:00 MST

This archive was generated by hypermail 2.4.0 : Wed Jun 21 2023 - 16:14:23 MST