Boron NMR Problems

From: <James.L.Simpson_at_invista.com>
Date: Fri, 4 Mar 2005 09:55:22 -0500

Greetings all,

I'm having a little trouble with 11B NMR. We have a Varian INOVA 500
system and I'm currently using a 10mm BB probe. Our sample of interest has
low levels of Boron (1-50ppm per species, using 1ml of Benzene-d6/3ml of
sample). I have been tuning the probe for each sample as it seems the
sample-to-sample variation is quite significant. The Boron species in the
sample are sensitive to the matrix (we do know that benzene does not
affect them). So, we've been using an external reference of 1 v%
BF3(Etherate) in benzene in a 5mm insert. We used the 1 v% BF3(etherate)
standard to optimize the pulse width for a full 90deg pulse (our samples
are too dilute to do any pw90 experiments). We are trying to optimize
acquisition time (at) and the delay (d1). Currently we've been varying the
at between 0.05 and 0.2 seconds with a delay anywhere from 0-0.3 seconds.
We are decoupling proton, but we've been unable to run a gamah2 on these
samples.

Our results are hit and miss, resolution pretty low. We've tried running
the sample w/o dilution, but we sacrifice lock. I'm not sure what else to
try. I even tried giving the magnet a Valentine's Day gift, but that
didn't seem to help either. Do our above parameters seem to be in the
ballpark for 11B NMR? Does anyone have any tips for running 11B NMR and/or
experience with low levels of Boron? Any help would be greatly
appreciated.


Jim Simpson
Invista
Orange, TX


This communication is for use by the intended recipient and contains information that may be privileged, confidential or copyrighted under applicable law. If you are not the intended recipient, you are hereby formally notified that any use, copying or distribution of this e-mail, in whole or in part, is strictly prohibited. Please notify the sender by return e-mail and delete this e-mail from your system. Unless explicitly and conspicuously designated as "E-Contract Intended", this e-mail does not constitute a contract offer, a contract amendment, or an acceptance of a contract offer. This e-mail does not constitute a consent to the use of sender's contact information for direct marketing purposes or for transfers of data to third parties.

Francais Deutsch Italiano Espanol Portugues Japanese Chinese Korean

www.invista.com/email_disclaimer.shtml




This communication is for use by the intended recipient and contains information that may be privileged, confidential or copyrighted under applicable law. If you are not the intended recipient, you are hereby formally notified that any use, copying or distribution of this e-mail, in whole or in part, is strictly prohibited. Please notify the sender by return e-mail and delete this e-mail from your system. Unless explicitly and conspicuously designated as "E-Contract Intended", this e-mail does not constitute a contract offer, a contract amendment, or an acceptance of a contract offer. This e-mail does not constitute a consent to the use of sender's contact information for direct marketing purposes or for transfers of data to third parties.

Francais Deutsch Italiano Espanol Portugues Japanese Chinese Korean

www.invista.com/email_disclaimer.shtml
Received on Fri Mar 04 2005 - 14:03:49 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 10 2023 - 17:13:18 MST