FnMode trouble

From: RAINER HAESSNER <Rainer.Haessner_at_ch.tum.de>
Date: Mon, 11 Mar 2002 12:16:48 +0000

Hi all,

recently we used older pulse programs (especially
HMQC and HMBC) with XWinNMR 3.0.
After data acquisition and processing nothing was correct
with the F1 axis.
We found FnMode as the source of the trouble.
If this new acquisition parameter is set, this becomes
used for processing too, MC2 is without of interest
in this case.
If you are doing experiments with pulse sequences which
doesn't use this new parameter, check out
     1s fnmode
prior to processing. This should be set to "undefined".

Regards

Rainer