Re: [AMMRL] Topspin 4.5.0 data transfer & backup for Windows 11

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Fri, 27 Jun 2025 17:31:28 +0000

Dear Piotr

Topspin and ICONNMR have built in archive functions.
In Topspin you can set the archive option in the preferences in the
acquisi tion section.

You can define the archive location here:

In Iconnmr it is done in the user management section under "other settings"

This is always a user specific setup.

Regards

Clemens



On 6/26/2025 7:08 PM, Piotr Dobrowolski via groups.io wrote:

Hello Everyone,

We just had a new Avance Neo 700 MHz instrument installed few days ago. It
is operated by Topspin 4.5.0 software running on a Windows 11 workstation.

I must admit I have very limited experience with Bruker spectrometers and
no experience with Topspin (any version) at all.

We would like to setup some kind of automatic data backup or data transfer
system for IconNMR user downloads. I could find some discussions about it in
the AMMRL archives but all related to Linux OS.

My question is: how to set it up in Windows 11? People recommended using OneDrive
but I don't really know how to do that.

For our two JEOL ECS-400 machines we use a NAS device connected to TTU network.
Maybe this is a better solution for our new AN700 backup...

My second question is related to that - these ECS machines are operated by
Windows 10 workstations and we create data backups using SyncToy tool. We will
be forced to upgrade to Windows 11 by the end of September. I heard there is
no SyncToy for that OS. How to setup such NAS backup without SyncToy in
Windows 11?

Thank you very much,

Piotr


--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2465): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2465__;!!PvDODwlR4mBZyAb0!RJVzQV2LSbsWqkFMKwWNWEqN_zj_60=
CcBkQr50zNssiLAYb3nBiEbtoO7VSDUJKcQbIjWSJCrb-qmhRqfs9rC474Krqa$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113853526=
/7559972__;!!PvDODwlR4mBZyAb0!RJVzQV2LSbsWqkFMKwWNWEqN_zj_60CcBkQr50zNssiLA=
Yb3nBiEbtoO7VSDUJKcQbIjWSJCrb-qmhRqfs9rC1DYb7E-$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






7UJkBSVhjFd5h8vA.png
(image/png attachment: 7UJkBSVhjFd5h8vA.png)

wun7FIHxFeXusxE3.png
(image/png attachment: wun7FIHxFeXusxE3.png)

Received on Fri Jun 27 2025 - 11:34:20 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST