Re: [AMMRL] Checking and editing magnet's base frequency on VNMRJ4 #agilent #varian #VarianAgilent

From: george gray via groups.io <george.gray.nmr=gmail.com_at_groups.io>
Date: Sun, 8 Jun 2025 06:22:20 -0700

Hi Emily,

Your sample (90% H2O) will have a very broad line at 600 MHz because of
"radiation damping". Try a "pure" D2O sample which should show a HDO
signal. It should have a linewidth appropriate for the shimming.

George Gray

Retired from Varian/Agilent

On 6/7/2025 2:03 PM, Shin, Ronald via groups.io wrote:
>
> Emily,
>
> This is from the good folks in Farmington, CT:
>
> https://health.uconn.edu/structural-biology/wp-content/uploads/sites/177/2017/10/Resetting_lock_freq_vnmrj.pdf
>
> I’d start by putting in an easy standard, such as one of your
> lineshape samples (99% d-acetone, 1% protonated chloroform), since
> this will have a very strong lock signal.
>
> Change the frequency downward in conservative increments, until you
> can see that strong lock signal (not a weak one like the sidebands),
> and try to get it to the middle of your numerical field.
>
> For Bruker spectrometers, the ‘cf’ command should access something
> similar where you can change the base frequency of the spectrometer.
>
> Ronald Shin
>
> University of Alabama at Birmingham
> O’Neal Comprehensive Cancer Center
> Central Alabama High Field Nuclear Magnetic Resonance Facility
>
> *From:*main_at_ammrl.groups.io On Behalf Of Emily Crull via groups.io
> *Sent:* Friday, June 6, 2025 11:44 AM
> *To:* main_at_ammrl.groups.io
> *Subject:* [AMMRL] Checking and editing magnet's base frequency on VNMRJ4 #agilent #varian #VarianAgilent
>
> Hi all,
>
> I'm working on a Unity Inova 600 MHz NMR. Recently, the console went
> offline due to a power issue. It has since come back online and we
> still have a magnet. However, it is experiencing a new problem.
>
> When acquiring a 1H spectra of our D2O standard (90% H2O, 10% D2O),
> the H2O peak is extremely broad (presumably due to poor shims) and
> appears at roughly 4.6 ppm. In addition to this, I cannot get a strong
> lock signal, even after sweeping all of the z0 range and phasing it.
> According to people who have been at WVU longer than I, it has had
> problems finding z0 automatically for years.
>
> I have the instructions for changing the lock frequency, but am
> wondering if we are also seeing the effects of magnet drift. *Is there
> a way to check the magnetic field and if necessary correct it in VNMRJ4?*
>
> Best,
>
> Emily
>
>

--
This email has been checked for viruses by Avast antivirus software.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2388): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2388__;!!PvDODwlR4mBZyAb0!TIovqGV3zxNqZvEZTBGtAfW9QL1zh3=
Mw-6ZY4rPBWYeY-b9Fbh02eXt315LMkLRm6-Ru1bfyul7qKex1Ou-MdWDFq_QK$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113521839=
/7559972__;!!PvDODwlR4mBZyAb0!TIovqGV3zxNqZvEZTBGtAfW9QL1zh3Mw-6ZY4rPBWYeY-=
b9Fbh02eXt315LMkLRm6-Ru1bfyul7qKex1Ou-MdV1J8DjQ$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!TIovqGV3zxNqZvEZTBGtAfW9QL1zh3Mw-6ZY=
4rPBWYeY-b9Fbh02eXt315LMkLRm6-Ru1bfyul7qKex1Ou-MdTi3s_Pr$
Mute #varian:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/varian__;!!PvDODwlR4mBZyAb0!TIovqGV3zxNqZvEZTBGtAfW9QL1zh3Mw-6ZY4r=
PBWYeY-b9Fbh02eXt315LMkLRm6-Ru1bfyul7qKex1Ou-MdR5ObtF9$
Mute #varianagilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/m=
ain/mutehashtag/varianagilent__;!!PvDODwlR4mBZyAb0!TIovqGV3zxNqZvEZTBGtAfW9=
QL1zh3Mw-6ZY4rPBWYeY-b9Fbh02eXt315LMkLRm6-Ru1bfyul7qKex1Ou-MdQuongXu$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Sun Jun 08 2025 - 07:09:49 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST