Dear Christian,

That is very helpful to know. The NMR lab was at a high temperature today
of 28 C I decided to wait until tomorrow to turn the spectrometers back
on. It should be much cooler by then.

Thank you! Best wishes,

Karl


On Thu, Jun 20, 2024 at 2:09 PM Christian Adam via groups.io wrote:

> Dear Karl,
>
>
>
> I experienced similar problems regularly during summer at my old lab. My
> experience is that at around 28°C / 83 F the first boards gave up and
> produced errors, though they did not suffer permanent damage and came back
> to life once the temperatures were lower. Regardless, I would encourage you
> to shut down the spectrometers before this point is reached.
>
> Bests,
>
> Christian
>
> *Dr. Christian Adam*
> Analytics and Material Physics
> Polymer, Quantitative and Heteronuclear NMR
>
> Phone: +49 621 60-58670, Mobile: +49 172 6352576, Fax: +49 621 66-58670,
> Email: christian.k.adam_at_basf.com
> Postal Address: BASF SE, RGA/AS - B009, Carl-Bosch-Strasse 38, 67056
> Ludwigshafen am Rhein, Deutschland
>
> BASF SE, Registered Office: 67056 Ludwigshafen, Germany
>
> Registration Court: Amtsgericht Ludwigshafen, Registration No.: HRB 6000
>
> Chairman of the Supervisory Board: Kurt Bock
>
> Board of Executive Directors:
>
> Markus Kamieth, Chairman;
>
> Dirk Elvermann, Michael Heinz, Anup Kothari, Stephan Kothrade, Katja
> Scharpwinkel
>
>
>

--

Karl D. Bishop, Ph.D.
Research Associate & Instrumentation Specialist
Department of Chemistry
146 Aubert Hall
University of Maine
Orono, ME 04469
207-581-1195
kbishop_at_maine.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1361): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1361__;!!PvDODwlR4mBZyAb0!R9BoFttNbgETImoYeZfHDg7ciREmZO=
u-4HMyKkHbZHuHB1wy2BxWgK3Ae18LvHlZzqjObDH9qcQQ_tIG$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106759058=
/7559972__;!!PvDODwlR4mBZyAb0!R9BoFttNbgETImoYeZfHDg7ciREmZOu-4HMyKkHbZHuHB=
1wy2BxWgK3Ae18LvHlZzqjObDH9qcMCKr68$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jun 21 2024 - 05:12:13 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST