Re: [AMMRL] Maximum room temperature for running NMR

From: Karel Klika <karkli_at_utu.fi>
Date: Wed, 19 Jun 2024 17:16:17 +0000

Dear all,

A high concentration of carbon dioxide is a very great danger irrespective
of the level of oxygen, I would strongly recommend against using dry ice in
this manner.

Regards,
K. Klika


________________________________________
> From: main_at_ammrl.groups.io on behalf of Mark Swanson
> Sent: 19 June 2024 19:40
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Maximum room temperature for running NMR

Hi Karl,

If you have a large tote of dry ice you could try blowing a fan over it and
see if that makes a difference. Or you could vent a liquid nitrogen tank.
Just make sure you have an oxygen sensor in the room. Our lab used to be
very warm in the mornings so I would always open all the doors to air it out
in the morning. I agree with Clemens about the humidity. One time in my
former life I gave an invited talk at the university of Rome. They took
me to see their MRI machines and it was rainy and muggy and they had the
windows wide open. I don’t think they even had air conditioning.
It was an interesting experience.

Cheers,

Mark

On Jun 19, 2024, at 7:18 AM, Clemens Anklin via groups.io <clemens=
=anklin.com_at_groups.io> wrote:

Hi Karl

our site specs define an upper limit of 25C (77 F). I do not think components
will die if you are a little bit above this temperature.
What you have to be concerned with is humidity. There are parts of the spectrometer
such as the heat sinks on the nitrogen stacks that are colder.

Humidity will condense on these surfaces.

If you have a cryoprobe or prodigy that body of the probe is also a bit colder
and you might see the surface get wet.

Regards

Clemens

On 6/18/2024 10:58 PM, kbishop_at_maine.edu<mailto:kbishop_at_maine.edu> wrote:
Dear AMMRL Colleagues,

In central Maine as in other parts of the Northeastern US we are experiencing
a heat wave. The central air conditioning for our building is down. We have
an INOVA 400 and Bruker Avance Neo 500 in our NMR lab. Four portable AC units
have been moved into the NMR lab to try to keep the lab cool. But they are
struggling to keep the room cool. I need to know at what temperature I should
shut down the NMR spectrometers.

Thank you.

Karl

Karl D. Bishop, Ph.D.
Research Associate & Instrumentation Specialist
Department of Chemistry
146 Aubert Hall
University of Maine
Orono, ME 04469
207-581-1195
kbishop_at_maine.edu<mailto:kbishop_at_maine.edu>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1353): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1353__;!!PvDODwlR4mBZyAb0!TKDgsf95vMjfCsVU56oi_ytJK3Dips=
1_ioeHAw_BOMmBKEvoYr7JD6Uway7tfUwYbPpTg_RwUtvnIA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106759058=
/7559972__;!!PvDODwlR4mBZyAb0!TKDgsf95vMjfCsVU56oi_ytJK3Dips1_ioeHAw_BOMmBK=
EvoYr7JD6Uway7tfUwYbPpTg_TjStMXCQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Wed Jun 19 2024 - 10:16:34 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST