From: Clemens Anklin <clemens_at_anklin.com>
Date: Wed, 19 Jun 2024 10:18:38 -0400

Hi Karl

our site specs define an upper limit of 25C (77 F). I do not think
components will die if you are a little bit above this temperature.
What you have to be concerned with is humidity. There are parts of the
spectrometer such as the heat sinks on the nitrogen stacks that are colder.

Humidity will condense on these surfaces.

If you have a cryoprobe or prodigy that body of the probe is also a bit
colder and you might see the surface get wet.

Regards

Clemens

On 6/18/2024 10:58 PM, kbishop_at_maine.edu wrote:
> Dear AMMRL Colleagues,
> In central Maine as in other parts of the Northeastern US we are
> experiencing a heat wave.  The central air conditioning for our
> building is down.  We have an INOVA 400 and Bruker Avance Neo 500 in
> our NMR lab. Four portable AC units have been moved into the NMR lab
> to try to keep the lab cool.  But they are struggling to keep the room
> cool.  I need to know at what temperature I should shut down the NMR
> spectrometers.
> Thank you.
> Karl
>
> Karl D. Bishop, Ph.D.
> Research Associate & Instrumentation Specialist
> Department of Chemistry
> 146 Aubert Hall
> University of Maine
> Orono, ME 04469
> 207-581-1195
> kbishop_at_maine.edu

--
==========================
=======================
Clemens Anklinemail:clemens_at_anklin.com
69 Chamberlain Road phone:978 692 1859
Westford MA 01886 fax :978 692 2826
http://www.anklin.com
$ twitter: _at_canklin
==========================
=======================


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1351): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1351__;!!PvDODwlR4mBZyAb0!QpgM2SyUm8s6UIsiVRbvOcjxAex45p=
Hm5LwA2Rj67uVe5kMx4JJpHMOnkOS8fK8MeL5iJDcp0kxht9s$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106759058=
/7559972__;!!PvDODwlR4mBZyAb0!QpgM2SyUm8s6UIsiVRbvOcjxAex45pHm5LwA2Rj67uVe5=
kMx4JJpHMOnkOS8fK8MeL5iJDcpg1mulT4$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jun 19 2024 - 07:18:49 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST