Hello Fu,
I don't have a "recent" HRMAS probe, but my newest one is from 2012 (850
MHz).
1H:Â 5.5 us _at_ 40 W
2H:Â 85 us _at_ 12 W
13C:Â 5 us _at_ 200 W
15N:Â 8.4 us _at_ 200 W
I hope this helps.
Bob.
On 2023-06-27 1:40 p.m., Fu Chen via groups.io wrote:
>
>
>
> Hi all,
> Can anyone share the parameters of π/2 observe pulse length and power
> level (1H, 31P, 13C and 15N) for Bruker 4mm HRMAS probe? The better
> the most recent probe.
> Thanks very much!
> Fu
> --
> Fu Chen, PhD
> Director of Analytical NMR Facility
> Department of Chemistry and Biochemistry
> B0208A Chemistry Building
> University of Maryland, College Park, MD 20742
> Tel: 301-405-1869
> fchen127_at_umd.edu
> https://blog.umd.edu/nmr/
>
--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.
McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#250): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/250__;!!PvDODwlR4mBZyAb0!XvS5VV_45RRAD1xxG6kvtKavTllam6Lh=
Hq37xwWL3NhfKtssnuQaIOWcdQtmRni6t9b5FCclOvd9YiEC$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/99814285/=
7559972__;!!PvDODwlR4mBZyAb0!XvS5VV_45RRAD1xxG6kvtKavTllam6LhHq37xwWL3NhfKt=
ssnuQaIOWcdQtmRni6t9b5FCclOtavVhDW$
Mute #bruker:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!XvS5VV_45RRAD1xxG6kvtKavTllam6LhHq37xw=
WL3NhfKtssnuQaIOWcdQtmRni6t9b5FCclOstprNZu$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Jun 28 2023 - 12:08:04 MST