Deuterium lock gone, proton channel NG

From: Lawrence Byrnes <numare_at_eclipse.net>
Date: Mon, 11 Jun 2001 11:58:46 -0400

Hello Spinlanders

Our system is now no longer functioning. I inserted a sample in CDCl3 this
morning with lock power set to 45. Initially, lock trace OK. After ca. 20
minutes, lock trace gone. Checked D3COD and DMSO-d6 as well. NG.

Proton line width went from normal to > 200 Hz. The resonance peaks look
like square waves.

Is it possible that the 10 MHz crystal oscillator in the synthesizer is no
longer stable? Is it possible to diagnose this problem with a 200 MHz
scope? We do not have a frequency counter here.

TIA, Lawrence


Lawrence Byrnes,
NMR Laboratory Manager,
Numare Spectralab Incorporated
10 Summit Avenue, # 6
Berkeley Heights, NJ, 07922
Research Director,
The Schoenenberg Instutute of Molecular Spectroscopy
75 Snyder Avenue, Berkeley Heights, NJ, 07922
Telephone 908-665-0066
Facsimile 908-464-4192
E-mail numare_at_eclipse.net
WWW http://www.eclipse.net/~numare/
Received on Mon Jun 11 2001 - 16:00:02 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 04 2023 - 15:46:27 MST