Hi All,
Many of the power suppies that we use in our instrument have TEMPERATURE
sensors!!! When the fan fails, the temperature rises and the temperature
sensor turns the power supply OFF. In many cases, replacing the fan ($20)
will fix the problem. Our Bruker Avance system even came with a replacement
fan in the tool kit!
Doug