Dear Karin,
- first you need to confirm "y" to the question to re-configure it
- next choose the enp2s0 network adapter (wlp3s0 sounds like a wireless connection)
- choose the 172.16.0.1 IP address
That should normally be it - the system should now try to check/update the
controller firmware after which you should get to an "Idle" state. If not try
to reboot the Linux computer after setacq.
If not done yet you may also still need to add the hostname (CSM-Agnes) to
the end of the first line in /etc/hosts before starting OVJ. Possibly rebuild
the database via "dbsetup" as vnmr1 user (has nothing to do with console comms
though).
Best regards
Bert
> Am 31.07.2026 um 22:29 schrieb Crowhurst, Karin A via groups.io:
>
> Hi everyone,
>
> I have made it further along in the attempted installation and setup of OpenVnmrJ
> to run my Agilent DD2. However, I cannot seem to get the computer to link with
> the NMR console. When I open the OpenVnmrJ software, it says the system is
“inactive” (even if I have used the “acqcomm” command in a terminal
> to start acquisition communications).
>
> To describe the computer system I am using, it is a Dell XPS, I have installed Ubuntu 24,
> and it has one built-in ethernet port, into which I have plugged the ethernet cable from
> the NMR, and an adapter ethernet port (plugged in via USB), to which the ethernet cable
> for the internet is attached. I have attached an image of what I see when I use the setacq
> command. It looks like the ethernet port for the NMR is not setup. Can anyone provide some
> insight into how I can make it communicate?
>
> Many thanks in advance,
> Karin Crowhurst
>
> Dr Karin Crowhurst
> Professor
> Graduate Coordinator
>
> Department of Chemistry and Biochemistry
> California State University Northridge
> 18111 Nordhoff St. MD 8262
> Northridge CA 91330-8262
> 818-677-4288
>
Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net Web: http://www.spin-doc.net
Representing:
Stelar Fast Field Cycling NMR Relaxometers (
http://www.stelar.it )
MestreLab NMR Software (
http://www.mestrelab.com
Barthel HF Amplifiers (
http://www.barthel-hf.de
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3630): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3630__;!!PvDODwlR4mBZyAb0!Uu_04UcnQAtQV0wXEghCT4aHccxIue=
GyD4xvQcqh_XEuh6Dz0gDfdudusEmKgcWg7q1F8AWeIgLqgm-fIg93xg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120541079=
/7559972__;!!PvDODwlR4mBZyAb0!Uu_04UcnQAtQV0wXEghCT4aHccxIueGyD4xvQcqh_XEuh=
6Dz0gDfdudusEmKgcWg7q1F8AWeIgLqgm_Bdwpxhg$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!Uu_04UcnQAtQV0wXE=
ghCT4aHccxIueGyD4xvQcqh_XEuh6Dz0gDfdudusEmKgcWg7q1F8AWeIgLqgm8E7Cig1Q$ [wa=
lt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Jul 31 2026 - 18:24:09 MST