Re: [AMMRL] 2H Inversion Recovery

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Thu, 30 Jul 2026 23:20:12 +0000

Hi Casey,

attached is a sequence draft. I have not tested it.

The long pulses limit your bandwidth. Luckily the common 2H bandwidth is small.
They will also become an issue if your relaxation times no longer are
significantly longer than the pulse length.

Regards

Clemens

On 7/30/2026 6:34 PM, Casey Simons via groups.io wrote:

Hey all,

I just had students request an inversion recovery experiment for 2H. I'm
currently running a Bruker BBO cyroprobe, so I'm using the sequence zg2h.
Has anyone altered this sequence to do an inversion recovery on 2H before
or does the long weak pulse affect the results?

Thanks!

Casey

Casey Simons
NMR Director
CAMCOR Nanofabrication and Imaging Facility
University of Oregon



--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3628): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3628__;!!PvDODwlR4mBZyAb0!QWCCBg8GNZlxB5NbOQbASudPDjYgdH=
ikYXC54ZO6QJfGIIR_-Inx3DE7zAx2dlhYCciXsQPXC0revPNtfuekoIkEu-cJ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120527503=
/7559972__;!!PvDODwlR4mBZyAb0!QWCCBg8GNZlxB5NbOQbASudPDjYgdHikYXC54ZO6QJfGI=
IR_-Inx3DE7zAx2dlhYCciXsQPXC0revPNtfuekoKn2jP_k$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






Received on Thu Jul 30 2026 - 16:57:50 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST