Hi Paris
the trick with 19F COSY experiments is having a long enough evolution time
in t1 to allow the magnetization to fully evolve. with 3J F,F couplings in
the range of 10 Hz you would get maximum transfer at ca 50 msec. So the COSY
LR help in building signal intensity. The next issue is that you have a very
wide sweep width and some small couplings. You need to acquire with sufficient
resolution in f2 to somwhat adequately digitze the signals. I would set
TD in F2 to 8K or larger. and TD in F1 to 256 or 512.
This should in principle give you a decent COSY.
Clemens
On 7/22/2026 2:17 PM, Paris Ning, PhD via groups.io wrote:
Hi everyone,
I have been trying to set up a 19F COSY experiment that can be used routinely
reliably. I am using PFOA as the model molecule and we have a Bruker system.
I've tried a few pulse sequences but none gave good results:
- COSY with gradient pulse for selection and purge pulse before d1 (cosygpppqf)
- COSY with delay for longrange coupling (cosylrqf) - tried with various delay
values from 0.1 ms to 50 ms
- Vanilla COSY
Wondering if anyone has a better COSY pulse sequence or another type of
correlation experiments to recommend.
Thanks a bunch!
Cheers,
Paris
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3613): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3613__;!!PvDODwlR4mBZyAb0!XAps7Dzqjc_5bY0_BzRsNvP5AnwPwP=
te5J5m3b8cFYVTqe61FZYzsyinaqsBdgDZ7k6bYRF1gM5ldJgvweuyA1RU5o5Q$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120399804=
/7559972__;!!PvDODwlR4mBZyAb0!XAps7Dzqjc_5bY0_BzRsNvP5AnwPwPte5J5m3b8cFYVTq=
e61FZYzsyinaqsBdgDZ7k6bYRF1gM5ldJgvweuyA0_4fqn5$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Jul 27 2026 - 13:47:44 MST