Hi Emily,
This happens when they have a lot of jobs (or jobs with lots of methods) in
their account. Delete the jobs and it’ll fix the problem, until he
has a lot of jobs again.
Tessa
On Jul 8, 2026, at 10:58 AM, Emily Crull via groups.io wrote:
Hi all,
One of the users of our JEOL 400MHz has this occasional bug where the sample
list will take upwards of 10 minutes to load in. The program itself boots
fine, it is literally just the sample list.
I can log in to my personal account and the console account and the list loads
basically instantly.
We can't seem to get this to occur reproducibly besides the fact that it only
happens on his account. He's not the newest account made and he has the same
permissions as everyone else.
The only thing he does that others don't is that he runs a lot of overnight
experiments, which means the computer goes into sleep mode with Delta open
on his account. The only solution we have found is to make a brand new account
for him. We did that once and now the account we created for him a few months
ago is experiencing the same issue.
Has anyone encountered this before and how did you fix it?
Many thanks,
Emily Crull
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3552): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3552__;!!PvDODwlR4mBZyAb0!WhtS_0qlngc1AGAqATRO1RwkNXArsJ=
i_6ly3jFnmdn77kr-i7G6828Ojz9TDFUb_-Gyl0eRUnE91Rx_QuBQP40DS9Ji7$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120174663=
/7559972__;!!PvDODwlR4mBZyAb0!WhtS_0qlngc1AGAqATRO1RwkNXArsJi_6ly3jFnmdn77k=
r-i7G6828Ojz9TDFUb_-Gyl0eRUnE91Rx_QuBQP46gHePVH$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Jul 08 2026 - 08:56:59 MST