Re: [AMMRL] J Young Tubes getting stuck in Bruker 400 with CASE autosampler

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Mon, 6 Jul 2026 13:44:11 +0000

HI Justin,

yes J-Young or LPV tubes are top heavy. Do you have a Kel-F spinner? those
have a but more mass and might work better.
the other options is lift tuning. I would have to look up the exact how to.
But maybe _at_Groves.Michael can join in.

I would suggest a faster decent and shorter hover time if possible.

Clemens


On 7/6/2026 9:00 AM, Justin Douglas via groups.io wrote:

Hello,

J Young tubes are getting stuck on occasion during the loading step in our
walkup Bruker 400 with sample case. I believe the issue is that these
top-heavy tubes are bouncing into a groove in the top of the upper barrel,
then wedging themselves diagonally between these grooves and the shuttle of
the autosampler.

I made a movie of the issue and have uploaded with annotations to youtube -
https://youtu.be/zpM9dHy0qWE

The system will try to reinsert 5 times before giving up. Many times one of
these attempts is successful. Sometimes the tube will not load after five
times and the systems goes into an error state, which requires manual
intervention to restart the queue. Even if the sample loads successfully
on the 2nd, 3rd, etc. try the extra time backs up the queue and seems to
irritate the students.

I would appreciate ideas to test my hypothesis (that tubes are jammed diago=
nally into between grooves on the BST and the sample shuttle) and ideas on =
how to address this issue. I can't imagine I am the first or only person t=
o have this issue.

A couple of additional comments:

These JYoung tubes were "shortened" by a glassblower to address the height
requirements of the sample case. Hence, it is at least conceivable that
tubes/valves have asymmetry. We have a tool to check tube straightness and
all of the tubes are acceptable. The tool doesn't check the valve, though.
Given that all tubes will load at least some of the time AND the amount of
money already spent on these J Young tubes, I would anticipate skepticism
(to put it politely) from the faculty if I suggest the issue is with the
tubes. In addition, there is another instrument right next to this 400 and
(to our knowledge) no JYoung tubes have ever gotten stuck in its sample case!
And I'll also note that (to our knowledge) no standard 5mm tubes from any
vendors (Wilmad, Norell, New Era, etc.) has ever gotten stuck in either
sample case.

One faculty member suggested trying to rig some sort of "bumper" that fits
on top of the shuttle and dampens the oscillations of the JYoung tube (but
presumably have no impact on normal tubes). Having spent the better part of
an afternoon perched on a ladder watching tubes go in and out of the instrument,
I think it is conceivable, but it seems like there is a lot that could go
wrong just above the bore of the magnet, which makes me nervous.

Thanks,

Justin


--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3535): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3535__;!!PvDODwlR4mBZyAb0!SfapxR8L_rGqvYTn5d2PSnDh4whZlg=
gn6VnbQp9igbx7HI7K3NqnwsXVK9JKCdEkVIHoWqHHKVoFe4sQap6nl5GT7_gs$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120139056=
/7559972__;!!PvDODwlR4mBZyAb0!SfapxR8L_rGqvYTn5d2PSnDh4whZlggn6VnbQp9igbx7H=
I7K3NqnwsXVK9JKCdEkVIHoWqHHKVoFe4sQap6nl9jBqIYz$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Jul 06 2026 - 06:48:38 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST