Dear Clemens,
The pump adapter is typically included with Varian NMR systems; however, I
haven't seen one provided with the Bruker systems.
We currently have three types of transfer lines from Bruker
Would it be possible to purchase a pump adapter directly from Bruker?
Best regards,
Weixing
> From: main_at_ammrl.groups.io On Behalf Of Anklin, Clemens via groups.io
> Sent: Wednesday, July 30, 2025 3:10 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Helium Transfer Lines - Refreshing vacuum. #poll-notice
Hi Ken
in general we recommend to pump it when it gets excessively cold during
transfers or when you think it takes a lot longer than usual to cool down.
This would indicate a soft vacuum.
So more need based than time based
Clemens
On 7/30/2025 12:08 PM, Kenneth Sharp-Knott via groups.io wrote:
A new poll has been created:
Inspired both by recent discussions on the forum about capturing helium
during a fill and a recent visit by an engineer to our facility.
How often do you pump down your helium transfer line?
I will admit that we never have since I joined the facility in 2013 and I
didn't realize that you could.
1. Never
2. Every fill
3. Every year or more frequently
4. More than yearly
5. What? You can pump down a helium transfer line???
Click Here To Vote <
https://ammrl.groups.io/g/main/vote?pollid=50667>
Do not reply to this message to vote in the poll. You can vote in polls only
through the group's website.
________________________________
Email Disclaimer: http://www.stjude.org/emaildisclaimer
Consultation Disclaimer: http://www.stjude.org/consultationdisclaimer
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2657): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2657__;!!PvDODwlR4mBZyAb0!US4Jzablbyf0sgE1N94HcxyK4nJl53=
oQpYhsbje2PXAgOjHM9niRMpZIY-eZxkDZACb3lDo0wFIL3QbryGlN80T0XEEz$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114451451=
/7559972__;!!PvDODwlR4mBZyAb0!US4Jzablbyf0sgE1N94HcxyK4nJl53oQpYhsbje2PXAgO=
jHM9niRMpZIY-eZxkDZACb3lDo0wFIL3QbryGlN8y1xYKxE$
Mute #poll-notice:
https://urldefense.com/v3/__https://ammrl.groups.io/g/mai=
n/mutehashtag/poll-notice__;!!PvDODwlR4mBZyAb0!US4Jzablbyf0sgE1N94HcxyK4nJl=
53oQpYhsbje2PXAgOjHM9niRMpZIY-eZxkDZACb3lDo0wFIL3QbryGlN81zjqQEf$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Aug 01 2025 - 08:27:42 MST