Re: [AMMRL] Helium Transfer Lines - Refreshing vacuum. #poll-notice

From: clarkson.michael via groups.io <clarkson.michael=gene.com_at_groups.io>
Date: Thu, 31 Jul 2025 11:09:57 -0700

On the principle that is rarely worth it to break a good vacuum, I would only
pump down a line once I have evidence that the outer jacket is cooling down
during the fill.

When I came to my current facility one of the transfer lines was very soft,
to the point where it simply would not cool down; this was not a cryofab
line so I had to by the manual actuator from a different company (the overall
design is the same as the unit Todd showed -- there is just a difference at
the attachment point, which may be a hex nut, KF, or something else depending
on manufacturer).

When this was happening it happened to be at the same time we were getting
a magnet install, so the technician was kind enough to give us an overnight
pull from his turbo. What kind of vacuum equipment and protocols are others
using for this task? Is everyone just using a standard vacuum pump? Has
anyone had experience wrapping the line in a heating mantle?

Sparky


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2647): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2647__;!!PvDODwlR4mBZyAb0!WiFeaJHSPPkfQNEAd-uNEO1mTAPiDU=
ZOj0avqSkOppQvjQMVvGjuhXZCamKykGlxaJJ2dSLsc-BUtk2vUWtBtOmBUF3g$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114451451=
/7559972__;!!PvDODwlR4mBZyAb0!WiFeaJHSPPkfQNEAd-uNEO1mTAPiDUZOj0avqSkOppQvj=
QMVvGjuhXZCamKykGlxaJJ2dSLsc-BUtk2vUWtBtIFQiY_K$
Mute #poll-notice:https://urldefense.com/v3/__https://ammrl.groups.io/g/mai=
n/mutehashtag/poll-notice__;!!PvDODwlR4mBZyAb0!WiFeaJHSPPkfQNEAd-uNEO1mTAPi=
DUZOj0avqSkOppQvjQMVvGjuhXZCamKykGlxaJJ2dSLsc-BUtk2vUWtBtFfFoRQ8$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jul 31 2025 - 11:38:20 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:12 MST