Hello Mike,
Congratulations on getting the console upgrade.
To enable "spooler", type "set" in the command line.
This opens the "set preferences" window. Near the top you will see
"Enable automatic command spooling"
Make sure that option is checked. Then click Apply at the bottom of the
window.
And so you should now have command spooling enabled.
PLEASE NOTE: This is a user specific change. If you have a multi user
environment, then this needs to be enabled for each user.
I haven't had any issues viewing old NMR data using the latest release
of TopSpin (even old DRX data recorded with XwinNMR).
However, if you want to acquire new data using exactly the same
parameters used to acquire an old dataset, then that won't work. I think
you will need to recreate the experiments to work with the new
hardware. It's not exactly from scratch since you know what should
work. But I think it's going to take some time. On the positive side,
it will give you a chance to really take the new hardware out for a
spin. And I think you will probably be able to create improved versions
of the old experiments.
You mention that you are running TopSpin 4.2
If that is the case, then before you invest any time with this, I
recommend upgrading to a newer version ASAP. To quote one of my
colleagues: "TopSpin 4.2 is the devil, and should never be used to drive
an NMR spectrometer!" The latest release is TopSpin v4.5. Though I
haven't yet installed that on any of our spectrometers, I am quite
certain it is better than v4.2.
Good luck,
Bob.
On 2025-07-24 10:39 a.m., Delk, Michael Joseph via groups.io wrote:
>
> Hello Everyone,
>
> We have been lucky enough to get our 700 upgraded to a new NEO console
> and therefore can finally run TS4.2. We have two items that have been
> noticeable right off the bat.
>
> One is the spooler is not immediately working. Does anyone recall the
> need to “turn this on”? Minor issue in the scheme of things but
> throwing it in this message.
>
> More importantly, the Bruker rep told us there are issues or
> compatibility problems with our older data from this magnet. This is
> old TS 2.6 data. Has anyone found conversion methods or other ways to
> still make this older work useful?
>
> As is common users here are more or less creating all the thing they
> do form previous setcs.
>
> Thanks in Advance,
>
> Mike
>
> Mike Delk
> NMR Manager
> University of Pittsburgh School of Medicine
> Department of Structural Biology
> 412.370.0066--
>
>
--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
--------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.
McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One
Spoon" wampum agreement.
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2555): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2555__;!!PvDODwlR4mBZyAb0!RsyY3xCB4YAnYJAJHbn6v7zKmgDGR9=
c64c75bkpYb8l1jgpYJLZcwQD8YOle1YL8b74nEPjIhiYMv50xOaPGa_maShKX$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114322054=
/7559972__;!!PvDODwlR4mBZyAb0!RsyY3xCB4YAnYJAJHbn6v7zKmgDGR9c64c75bkpYb8l1j=
gpYJLZcwQD8YOle1YL8b74nEPjIhiYMv50xOaPGa4FnrxsT$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Jul 24 2025 - 10:48:52 MST