Re: [AMMRL] Topspin 4.5.0 (NEO700) topshim problem in ICONNMR

From: Piotr Dobrowolski via groups.io <piotr.dobrowolski=ttu.edu_at_groups.io>
Date: Thu, 17 Jul 2025 11:04:47 -0700

Hi,

It turned out that IconNMR was loading an outdated shim file as "good" starting
shims. I was able to fix that in Configuration - "Solvent/Probe Dependencies".
Now it loads my last 3D Topshim (sucrose standard) result and Topshim works
very well and fast in IconNMR.

Thank you Bob and Thomas for your suggestions.

Piotr


-=-=-=-=-=-=-=-=-=-=-=-