I’d check your Lock/Shim Options to ensure you have the correct Shim
Program. Double check that it shims every experiment. Also, go into the
Solvent/Probe Dependencies submenu and check that you have a good default
shim file associated with your solvents and that shim file loading in enabled.

Dan

-=-=-=-=-=-=-=-=-=-=-=-