Re: [AMMRL] No Na23 Tune dip on 500 MHz BBFO probe (Neo500) [Solved]

From: Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Date: Wed, 16 Jul 2025 05:48:50 -0700

Thanks everyone.

I ended up using ATMA to 13C as suggested.  And then using ATMM in 13C
I movied the dip to 132 and wrote those values down.

Then using ATMM with Na23, I set the coarse values I just found and was able
to find the dip. I'm sort of surprised I was never able to find the dip
otherwise.

Case closed. Thanks again.


-=-=-=-=-=-=-=-=-=-=-=-