From: Bob Berno <bberno_at_mcmaster.ca>
Date: Mon, 15 Jul 2024 13:19:08 -0400

Hello again TJ,

Regarding HASKRIS model number:  I wrote down the wrong number in the
earlier message.
Model:  INDOOR  WW2-230V-ABHK(R4)-CSA

Cheers,
Bob.
On 2024-07-15 12:13 p.m., Bob Berno wrote:
> Hello TJ,
>
> I hope you can sort out the problem.
> On a related note, if point 3(b) - that your UPS can't handle the
> power spike when pumps/compressors power on - is the root cause of
> your problem, then when the UPS fails, it may automatically switch to
> "bypass" mode - which I mentioned in point 3 (c).
>
> Regarding the automatic switch to city water:  that was a feature in
> the particular model of HASKRIS chiller we ordered.  I had asked about
> backup city water. They said that if we could but the necessary
> plumbing in pace, they could sell us a chiller that could
> automatically switch between chilled water and city water.
> NOTE:  once the system switches to city water, one has to manually
> switch back to chilled water.  I think that is a safety feature, which
> makes sense.
>
> For your reference, our HASKRIS chiller model is: INDOOR WW2-230V-10-60Hz
>
> Good luck,
> Bob.
>
> On 2024-07-15 11:33 a.m., Thusitha Jayasundera via groups.io wrote:
>>
>>
>> Thanks Bob--I know that we definitely do not have a city water
>> backup. I wasn't even aware that this was a possibility, and should
>> check on this immediately. Thus far, the issue has always occurred
>> during a power switch, and I think the point you make in 3 (b) is
>> likely where we went wrong--so I should look into that as well.
>>
>> Much appreciation to both of you for your suggestions!
>>
>> Best,
>>
>> TJ.
>>
>>
>
> --
> Dr. Bob Berno
> Manager, NMR Facility
> McMaster University, ABB-156
> 1280 Main St. West
> Hamilton, ON L8S 4M1
> T: 905-525-9140 x24158 or x24686
> ------------------------------------------------------------------------
> This email contains CONFIDENTIAL information intended for the specific recipient(s).
> If you have an accommodation need for a planned meeting, please e-mail me directly and
> I will do my best to make appropriate arrangements.
> Should you require any materials sent via this e-mail address in an alternate/accessible
> format, please let me know.
>
> McMaster University sits on the traditional Territories of the Mississauga and
> Haudenosaunee Nations, and within the lands protected by the "Dish With One
> Spoon" wampum agreement.
>
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
--------------------------------------------------------------------------

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1436): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1436__;!!PvDODwlR4mBZyAb0!XqjOXM5LXkrs9J7Xl9G-docgSw8jyD=
VT-lt7t_6nOAJYSSxZWTT75os3b-9m3sxNMeGhs5fW7tgWVg-P$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/107189604=
/7559972__;!!PvDODwlR4mBZyAb0!XqjOXM5LXkrs9J7Xl9G-docgSw8jyDVT-lt7t_6nOAJYS=
SxZWTT75os3b-9m3sxNMeGhs5fW7q71QOgo$
Mute #ups:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/muteha=
shtag/ups__;!!PvDODwlR4mBZyAb0!XqjOXM5LXkrs9J7Xl9G-docgSw8jyDVT-lt7t_6nOAJY=
SSxZWTT75os3b-9m3sxNMeGhs5fW7tYs2SHv$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!XqjOXM5LXkrs9J7Xl9G-docgSw8jyDVT-l=
t7t_6nOAJYSSxZWTT75os3b-9m3sxNMeGhs5fW7k1qttSb$
Mute #haskris:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/haskris__;!!PvDODwlR4mBZyAb0!XqjOXM5LXkrs9J7Xl9G-docgSw8jyDVT-lt7=
t_6nOAJYSSxZWTT75os3b-9m3sxNMeGhs5fW7v13d6Ca$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Jul 15 2024 - 10:19:14 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 03 2024 - 15:02:58 MST