Re: [AMMRL] Sieves for drying NMR solvents?

From: Akien, Geoffrey <g.akien_at_lancaster.ac.uk>
Date: Fri, 12 Jul 2024 20:57:15 +0000

Hi Sameer,

Acetone is one of the few solvents that do not respond well to molecular sieves
of any type: https://pubs.acs.org/doi/10.1021/jo00414a038 (see Table IV)

The end result is self-condensation and an increase in water content - you'd
be better off not using any at all! Can't say I've tried it myself, but
apparently B2O3 is the trick with this particular case.

Geoff